Hydrologic

Electrical conductivity transmitter with an inductive probe


TRANSMITTERS Alphée 3100/3200
INDUCTIVE PROBES
Alphée 3120/3220


  • Measurement of uncompensated of a liquid using the inductive effect.
  • Measurement of liquid temperature using a PT100 incorporated into the probe.
  • The transmitter can be switched to standby by an external control signal (for low current drain).
  • 2 measurement ranges :
    • 0 - 7 ms/cm
    • 0 - 420 ms/sm
  • Can be used with series 2000 data acquisition units.
  • Housed in a steel box for mounting on a DIN rail or in a plastic enclosure for wall mounting.
  • The probe is moulded with sensitive components which are not in contact with the liquid; this gives maximum strength and reduced clogging.
  • Resists chemical attack.
Technical
specifications
 

TRANSMITTER  
Range of uncorrected
conductivity measurement :
Alphée 3100 : 0-420 ms/cm
Alphée 3200 : 0-7 ms/cm
Range of temperature measurement -30°C à +70°C
Power supply 12 VDC (9 à 36 VDC)
Average current consumption 160 mA during measurement;
300 µA in standby
Temperature range -20°C à +60°C
Relative humidity < 80% without condensation

HOUSED IN PAINTED METAL BOX FOR DIN RAIL
Dimensions 200 x 122 x 43 mm
Protection index IP 40
IP 65 in plastic housing for wall mounting

CONDUCTIVITY ANALOGUE OUTPUT
Current
4 - 20 mA ( 0 to 500 ohms,
in accordance with standard)
Accuracy
< 1% of measurement range

TEMPERATURE ANALOGUE OUTPUT
Current
4 - 20 mA ( 0 to 500 ohms,
in accordance with standard)
Accuracy
< 1% of measurement range
Connections
screw terminal strip

PROBE
Range of uncorrected
conductivity measurement :
Alphée 3120 : 0-420 ms/cm
Alphée 3220 : 0-7 ms/cm
Range of temperature measurement
using incorporated PT100

-30°C à +70°C
Connections via 16 way connector
IP 68 water tightness (8 bars)
7 pair multi-core cable 0,22 mm2 + screen
Overall dimensions 164 x 44 x 43 mm
Can be immersed up to a depth of 70 m
Chemical resistance good resistance to hydrolysis,
sea water, acid and diluted bases.

STANDARDS/DIRECTIVES
CE MARK
 
CEM C46-023 ou 801-4
Level 3 protection tests